General Purpose Triaxial Vibration Transducers AP38, AP38P

    Destination:  

  • Simultaneous measuring of triaxial-oriented vibration and shock accelerations in portative diagnostic systems and during laboratory studies.
    Features:  

  • Three shear sensing elements design.
  • AP38 has fixed three-core low-noise cable, AP38P has built-in connector.
  • Combination of small overall dimensions and acceptable sensitivity.
Parameter Unit АР38*3 АР38P
Axial sensitivity (± 20 %) pC/g*1 10
Relative transverse sensitivity % < 5
Amplitude range g*1 ± 5 000
Max.shock limit (peak value) g*1 ± 10 000
Operating temperature range °C - 60…+ 150
Frequency range (ripple ± 1dB) Hz 0.5…10 000
Self-resonant frequency in attached condition kHz > 35
Strain sensitivity g·m/µm < 0.02
Capacitance pF 900…1 200
Insulation resistance in normal conditions MOhm > 10 000
Housing material - titanium alloy
(stainless steel)*2
Built-in cable length m 2*2 -
Weight (without cable) gram 26 (38) 18 (28)
Suplied accessories - screw М4×20> screw М4×20,
cable AK03-2m*2

General view of AP38

General view of AP38P

The electrical circuit of AP38

The electrical circuit of AP38P


*1 – 1g = 9.807 m·s-2 or 10 m·s-2 = 1.02 g;

*2 – to be determined when making out an order;
*3 – has been entered in State list of measuring means.