- Destination:
- Measurement of vibration parameters (vibration acceleration)in the regime of charge amplifier.
- Features:
- Acceleration measuring.
- Rating mode.
- Overshoot indication mode.
- Low noise level.
- Operation control mode through USB-port.
- Housing 19″-metal, height – 2U, depth – 260mm.
| Parameter | Unit | AP5220-8 |
|---|---|---|
| Input charge range | pC | 105 |
| Frequency range at the level -1 dB | Hz | 0,02 … 100 000 |
| Input resistance | Om | > 109 |
| Output resistance | Om | < 100 |
| Maximum output voltage at the noon-linear distortion factor < 5 % | V | ± 10 |
| RMS noise value (reduced to the input) | µV | ≤ 5 |
| Voltage ranging factor, (error ± 0,5 %) |
mV/pC | 0,1; 0,2; 0,5; 1; 2; 5; 10; 20; 50; 100 |
| КInfluence coefficient of ambient air temperature change within the range from 40ºС on amplification and conversion factor |
%/ºС | ≤ 0,025 |
| Conditioning factor current | pC/Eg. | 1 … 9,99 with step 0,01 |
| Built-in high-frequency filters (HFF) with the AFC decay no less than 40 dB/decade at the level –1dB: | Hz | 0,02; 0,2; 1; 2; 10; 100 |
| Built-in low-frequency filters (LFF) with the AFC decay no less than 40 dB/decade at the level –1dB | kHz | 1; 10; 30; 100 |
| Supply voltage | V | + (12 ± 1) |
| Consumption current | А | < 1 |
| Input connector type | - | BNC, 10-32 UNF |
| Output connector type | - | BNC |
| Power supply | - | Network from adapter |
| Dimensions | mm | 481×320×135 |
| Weight | gram | 2 000 |
- Designed for a work:
- with vibration transducers АР10, АР11, AP12, АР19, АР20, АР21, АР22, АР23, AP24, АР28, АР28B, АР30, АР31, АР32, АР33, АР34, АР35, АР37, АР38, АР39, АР40, АР46, АР47, АР48, АР49, АР50, АР57, AP58, АР77, АР78, АР79, АР80, АР81, АР85,АР90,АР91;
- with force transducers АС20, АС21, АС22, AC23,АС26;
- with dynamic pressure transducers PS01,PS02;
- with impulse hammers AU01.