AP2098-30-02
Purpose:
- Measuring vibration and shock acceleration in diagnostic systems and laboratory tests under strong electromagnetic fields.
Features:
- High axial sensitivity, self-resonant frequency and shock resistance.
- Low sensitivity to electromagnetic fields.
- Built-in preamplifier.
- Wide range of power supply voltage and current.
- Low level of intrinsic noise.
-
ParameterUnits of MeasureAP2098-30-02
-
Conversion factormV/ms23
-
Relative coefficient of transverse conversion%<5
-
Maximum value of measured acceleration amplitudem/s21 600
-
Maximum shock limit (peak value)g±1 500
-
Operating temperature range°С–40...+125
-
Operating frequency range (ripple ± 1 dB)Hz0.5...12 000
-
Self-resonant frequency in the fastened statekHz>40
-
Noise level, RMS (1 Hz ÷ 10 kHz)m/s2<0.003
-
Output resistanceΩ<500
-
Power supply: - voltage; - currentV; mA+(15...30); 2...20
-
Constant output voltage levelV8...13
-
Temperature influence coefficient%/°С±0.2
-
Housing material-stainless steel
-
Connector type-AR03(10-32UNF)/cable outlet
-
Weight (without cable)kg0.040
-
Supplied accessories-cable AK15 stud AH0105
-
Options-T, N
Optional accessories:
- charge and voltage amplifiers AP5100, AP5200, AP5210, AP5230, AP5230-1, AP5240B1, AP5240B1T, AP5240A1B1;
- power units AS01, AS07;
- matching devices AG01, AG02;
- cable AK26;
- cable adapters AR01, AR04;
- magnet mounting bases AM01, AM03, AM05, AM08;
- calibrators AT01m, AT02, AT04;
- stud AH0805.