AP2039-10

Purpose:

  • Measuring vibration and shock acceleration in diagnostic systems and laboratory tests under strong electromagnetic fields.

Features:
  • Built-in preamplifier.
  • Electrical insulation of sensitive elements from the housing.
  • Wide range of power supply voltage and current.
  • Low level of intrinsic noise.
  • High axial sensitivity, self-resonant frequency and shock resistance.
  • Low sensitivity to electromagnetic fields.
  • Parameter
    Units of Measure
    AP2039-10
  • Conversion factor
    mV/ms2
    1
  • Relative coefficient of transverse conversion
    %
    <5
  • Maximum value of measured vibroacceleration amplitude
    m/s2
    5 000
  • Maximum shock limit (peak value)
    g*1
    ±3 000
  • Operating temperature range
    °С
    –40...+125
  • Operating frequency range (ripple ± 1 dB)
    Hz
    0.5...5 000
  • Adjustment resonance frequency in axial direction
    kHz
    >15
  • Noise level, RMS (1 Hz ÷ 10 kHz)
    m/s2
    <0.01
  • Output resistance
    Ohm
    <500
  • Power: - voltage; - current
    V; mA
    +(18...30); 2...20
  • Constant output voltage level
    V
    8...13
  • Temperature influence coefficient
    %/°С
    ±0.2
  • Run mode setting time
    s
    4
  • Housing material
    -
    titanium alloy
  • Connector type
    -
    AR1601 (4-pin, 1/4-28UNF)
  • Weight (without cable)
    kg
    0.0145
  • Supplied accessories
    -
    cable AK41E1В3, screw M5 × 16
External view of АР2039-10
Electrical circuit of АР2039-10

Optional accessories:

  • charge and voltage amplifiers AP5110, AP5200, AP5210, AP5230, AP5230-1, АР5240В1, АР5240В1Т, АР5240А1В1;
  • power units AS01, AS07;
  • matching devices AG01, AG01-3, AG02, AG02-3, AG03;
  • calibrators AT01m, AT02, AT04;
  • cables AK58;
  • magnet mounting bases AM01, AM03, AM05, AM08.

*1 – 1g = 9.807 m·s-2 or 10 m·s-2 = 1.02 g.