AP2034-10

Purpose:

  • Measuring vibration and shock acceleration in diagnostic systems and laboratory tests under strong electromagnetic fields.

Features:
  • High axial sensitivity, self-resonant frequency and shock resistance.
  • Low sensitivity to electromagnetic fields.
  • Built-in preamplifier.
  • Wide range of power supply voltage and current.
  • Low level of intrinsic noise.
  • Parameter
    Units of Measure
    AP2034-10
  • Conversion factor
    mV/m·s2
    1
  • Relative coefficient of transverse conversion
    %
    <5
  • Maximum value of measured acceleration amplitude
    m/s2
    5 000
  • Maximum shock limit (peak value)
    g
    ±3 000
  • Operating temperatrue range
    °С
    –40 ... +125
  • Operating frequency range (ripple ± 1 dB)
    Hz
    0,5 ... 20  000
  • Adjustment resonance frequency in axial direction
    kHz
    60
  • Noise level, RMS (1 Hz ÷ 10 kHz)
    m/s2
    <0,015
  • Output resistance
    Ω
    <500
  • Power: - voltage; - current
    V; mА
    +(18 ... 30); 2 ... 20
  • Constant output voltage level
    V
    8 ... 13
  • Temperature influence coefficient
    %/°С
    ±0,2
  • Housing material
    stainless steel (titanium alloy)*2
  • Built-in cable length
    m
  • Connector type
    AR02 (M3)
  • Weight (without cable)
    п
    4
  • Supplied accessories
    cable АК06, stud АН0103
External view of AP2034-10
Electric circuit of AP2034-10
Wiring diagram for connecting to dataacquisition equipment

Optional accessories:

  • calibrators AT01m, AT02;
  • charge and voltage amplifier АР5240А1В1;
  • adapter boxes AG01, AG02;
  • wax mastic AW01;
  • stud АН0503, АН1003.


    *1 – 1g = 9,807 m·s-2 or 10 m·s-2 = 1,02g ;
    *2 – to be determined at customer’s request;
    *3 – registered in the State Register of measuring instruments.

Catalog